EXPLORE
CLINICAL ASPECTS

Clinical aspects

The following graphs are included to explain in detail the benefits of the Oxygen Therapy by Invacare. Experts and patients can find here the comparisons between the continuous and the pulse dose delivery systems.

Significant
waste of Oxygen

  • More than 75% of O2 flow is wasted during exhalation and in dead space.
  • At 2 LPM, a patient only uses - 0.5 LPM (22ml x 20).
oxygen concentrator machine

RR 20/min; I:E 1:2, inspiration 1s, 2LPM = 33ml/s

Pulse Dose

Clinically Efficient:: An oxygen bolus is delivered at the beginning of the inspiration, optimizing gas exchange.

Fixed Minute Volume:: SensiPulse technology adjusts the bolus volume per breath in response to the patient�s respiratory rate to ensure a constant minute volume of O2 per setting.

Clinically efficient oxygen delivery
The Sensi-Pulse conserver delivers only clinically beneficial oxygen at the beginning of inspiration allowing the maximum opportunity for gas exchange.

Pulse Dose in PLATINUM MOBILE

oxygen concentrator moldova

Bolus and MinuteVolume

PLATINUM MOBILE

Trigger Sensitivity: 0,18 cmH2O

RR 20/min

Setting Bolus Size Minute Volume
1 11 ml 220 ml/min
2 22 ml 440 ml/min
3 33 ml 660 ml/min
4 44 ml 880 ml/min

+
Pulse Dose in SOLO2

home oxygen concentrator

Bolus and MinuteVolume

SOLO2

Pulse Dose in SOLO

RR 20/min

Setting Bolus Size Minute Volume
1 18 ml 360 ml/min
2 36 ml 720 ml/min
3 54 ml 1080 ml/min
4 72 ml 1440 ml/min
5 90 ml 1800 ml/min

Pulse Dose in HomeFill

medical oxygen tank

HomeFill Refillable
Oxygen Cylinders

CONSERVER CYLINDERS Duration IN HOURS
Conserver setting 1 2 3 4 5 Continuous flow (2L/min) Full time
1 Litre Cylinder 9.3 5.9 4.5 3.6 3.0 1h 20min. 1h 20min.
1.7 Litre Cylinder 14.3 9.1 6.9 5.6 4.7 2LPM 5min. 1h 40min.
Setting Bolus Size Minute Volume
1 18 ml 360 ml/min
2 36 ml 720 ml/min
3 54 ml 1080 ml/min
4 72 ml 1440 ml/min
5 90 ml 1800 ml/min